| High voltage |
|
| HV adjustment range: |
10 - 160 kV |
| HV adjustment increments: |
0,1 kV/step |
| Accuracy: |
± 1 % (demand value) |
| Reproducibility: |
better ± 0,1 % |
| HV ripple (with 20 m cable): |
5 V/mA, min. 20 V |
| Temperature induced drift: |
< 50 ppm/°C |
|
|
| Tube current |
|
| Adjustment range: |
0 - 10 mA |
| Adjustment increments: |
0,1 mA/step |
| Tube current stability: |
± 0,1 % (demand value) |
|
|
| Exposure timer |
|
| Runtime: |
max. 99 minutes |
| Input: |
via keyboard/menu |
| Display: |
via display (LED) |
|
|
| Programmed operation: |
|
| Memory capacity: |
99 |
| Programming: |
via keyboard |
| Tube conditioning: |
5 programmes |
| Interface(s): |
2 x RS232 |
|
|
| Pre-warning time |
|
| Continuously adjustable from 0 to 99 seconds,
tunable in 1-second-steps via menu navigation and keyboard. |
|
|
Environmental condition
(Protection class IP44) |
|
| Duty cycle: |
< 1000 W - unlimited
> 1000 W - max. 30 min. |
| Operating temperature: |
0 to 40°C |
| Storage temperature: |
-25 to +70°C, relative humidity 80 %, non-convective air |
|
|
| Focal spot selection |
|
| Focal spot selection takes place at the push
of a button on the keyboard (2 symbols existing). Switching over only is
possible while HV is turned off. |
|
|
| Safety provisions |
|
| Two independent monitored safety circuits (fail-safe,
24 V). Continuous monitoring with automated shut-down and real text
failure indication on display. Automatic discharging device at shut-down.
Fail-safe monitoring of an external warning lamp and an additional
warning lamp in the remote control (option). External warning lamp is
based on LED technology. Two additional potential-free contacts for
start-up interlock. Additional emergency stop buttons can be connected. |
|
|
| Mains supply |
|
| 230 VAC, 16 A, +6 %, -10 %, Protection 16 A
at 1,6 kW |
|
|
| Mechanical data: |
|
| Dimensions (H x D x W): |
1400 x 870 x 610 mm |
| Weight (with 20 m HV cable): |
170 kg |
|
|
|

|